1. A computational study of artery curvature and endograft oversize influence on seal zone behavior in endovascular aortic repairŽiga Donik, Willa Li, Blessing Nnate, Joseph A. Pugar, Nhung Nguyen, Ross Milner, Enrique Cerda, Luka Pocivavsek, Janez Kramberger, 2024, izvirni znanstveni članek Opis: Thoracic endovascular aortic repair (TEVAR) is a minimally invasive procedure involving the placement of
an endograft inside the dissection or an aneurysm to direct blood flow and prevent rupture. A significant
challenge in endovascular surgery is the geometrical mismatch between the endograft and the artery, which
can lead to endoleak formation, a condition where blood leaks between the endograft and the vessel wall. This
study uses computational modeling to investigate the effects of artery curvature and endograft oversizing,
the selection of an endograft with a larger diameter than the artery, on endoleak creation. Finite element
analysis is employed to simulate the deployment of endografts in arteries with varying curvature and diameter.
Numerical simulations are conducted to assess the seal zone and to quantify the potential endoleak volume
as a function of curvature and oversizing. A theoretical framework is developed to explain the mechanisms of
endoleak formation along with proof-of-concept experiments. Two main mechanisms of endoleak creation are
identified: local buckling due to diameter mismatch and global buckling due to centerline curvature mismatch.
Local buckling, characterized by excess graft material buckling and wrinkle formation, increases with higher
levels of oversizing, leading to a larger potential endoleak volume. Global buckling, where the endograft
bends or deforms to conform to the centerline curvature of the artery, is observed to require a certain degree
of oversizing to bridge the curvature mismatch. This study highlights the importance of considering both
curvature and diameter mismatch in the design and clinical use of endografts. Understanding the mechanisms
of endoleak formation can provide valuable insights for optimizing endograft design and surgical planning,
leading to improved clinical outcomes in endovascular aortic procedures. Ključne besede: finite element analysis (FEA), endovascular aortic repair, endoleak, stent graft, TEVAR, EVAR Objavljeno v DKUM: 29.07.2024; Ogledov: 86; Prenosov: 39
Celotno besedilo (2,83 MB) Gradivo ima več datotek! Več... |